Choosing and caring for NiTi Perio Saw® tips

The NiTi Perio Saw® is supplied as four tip shapes, individually or as a set of four. This page collects, in one place, how the instructions for use (IFU) match each tip to a tooth position and how the instrument is reprocessed between procedures. It is written for the whole surgical team – in a veterinary practice that includes the veterinary nurses and technicians who prepare and reprocess instruments, and in a dental practice the dental nurses and assistants who do the same. The reprocessing conditions below are quoted from the veterinary edition of the IFU; the human dental edition has its own IFU, which is supplied with that instrument.

Four tips by tooth position

Tip Model code Intended position (IFU) Working tip length
Straight NiTi Periosaw-ST-02 Anterior teeth 36 mm
Curved NiTi Periosaw-CD-02 Canine or premolar 36 mm
Angled NiTi Periosaw-AD-02 Mesial and distal 42 mm
Flat angled NiTi Periosaw-FA-02 Interproximal 42 mm

All tips share a 100 mm stainless-steel handle (SUS304, ASTM A276) and a nickel-titanium shape-memory alloy tip (ASTM F2063) with a micro-serrated edge [1]. The suffix -02 denotes the 2 mm tip width [1]. As with other extraction instruments, the instrument should suit the root it is used on [2].

Before each use

The IFU checklist [1]: confirm that the patient has no known reaction to, or incompatibility with, nickel or titanium; check the package for breakage; inspect the instrument for damage, breakage, deformation or foreign material; sterilise it as below; and read the IFU. Do not use a damaged, bent or contaminated instrument. The recommended operating-room temperature is 15–30 °C [1].

Sterilisation

The instrument is supplied non-sterile and must be sterilised before first use and before every later use [1].

Method Configuration Temperature Exposure Drying
Autoclave Wrapped (sterilisation pouch or wrap) 134 °C 15 minutes 20 minutes

Keep it in the pouch or wrap until use [1].

Cleaning after use

Remove blood and debris under running water with a cleaning brush; clean with an instrument washer and instrument detergent following their manufacturers' instructions; dry with a clean cloth or a dryer; inspect again for damage, deformation and foreign material; then pack and sterilise [1]. General veterinary dental guidance likewise expects dental instruments to be cleaned, inspected and steam-sterilised in packs [3].

Storage and retirement

Store at 15–30 °C and 1–75 % humidity, away from sunlight, with nothing loaded on top of the package [1]. The IFU gives no sharpening procedure: the micro-serrated nickel-titanium edge is not resharpened. Retire a tip that is worn, damaged or deformed and dispose of it as sharp metal clinical waste under local rules for clinical waste [1].

Who does what in the clinic

The veterinary IFU limits use to veterinarians and trained veterinary staff [1]; the human dental edition is for dentists. Who may perform extractions differs by country – in the United Kingdom, for example, see the RCVS guidance on delegation to veterinary nurses [4] – so nurses, technicians and assistants should follow their national rules. Preparing the pack, reprocessing and inspection are typically shared by the team; this guide covers those tasks, not extraction technique. For technique, see Cutting the periodontal ligament in dog and cat extractions.

Sources

  1. HANS Korea Co., Ltd. Instructions for Use – NiTi Perio Saw (Veterinary), HKIFU_VT_001 Rev. 0, 30 September 2026. PDF
  2. Lobprise H. Essential tools to have in a dental extraction kit, according to a dental specialist. dvm360, 23 December 2025. https://www.dvm360.com/view/essential-tools-to-have-in-a-dental-extraction-kit-according-to-a-dental-specialist
  3. Altier B. Dentistry instruments and equipment: uses and maintenance. Today's Veterinary Nurse (updated November 2021). https://todaysveterinarynurse.com/dentistry/dentistry-instruments-and-equipment-uses-and-maintenance/
  4. Royal College of Veterinary Surgeons. Delegation to veterinary nurses (Code of Professional Conduct, supporting guidance 18). https://www.rcvs.org.uk/veterinary-professionals/conduct-and-guidance/supporting-guidance/18-delegation-to-veterinary-nurses

Reviewed: 22 October 2026

Back to blog